| MPLAB ICD 4 Help |
|
|
If the MPLAB X IDE or MPLAB IPE cannot communicate with the debugger, the debugger may need to be forced into bootload mode (download new firmware). Some possible reasons could be the following:
• If steps 1-5 in “General” did not correct the debugger issue.
• Refer to “Indicator Lights (LEDs)” for more information on LED modes and bootloader errors.Preferred Method: If you have the recommended Microchip 9V power supply (AC002014), plug it into the MPLAB ICD 4, but unplug the power cord from the wall or power strip.Alternative Method: If the recommended power supply is not being used, make sure no power is connected to the debugger. Then, insert a small metal screwdriver into the 9V female barrel connector so that it contacts the center pin and the metal tang at the bottom of the jack.Figure: Alternative Method of Creating a Short![]()
4. As soon as both LEDs flash quickly, immediately unplug the power cable from the debugger or remove the screwdriver (within 1 second). If the LEDs are steady on purple, proceed to the next step. If the LEDs are blue, you did not perform the action fast enough and must repeat the process starting at step 1).
|
|